Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Transmation 1074: (1 year) CAL VER /5520 DATE: 11-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 8 NUMBER OF LINES: 94 CONFIGURATION: Fluke 5520A CONFIGURATION: HP 34401A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Connect UUT to wall power adaptor and turn unit ON. 1.004 DISP 1.004 DISP Warm-up time: 30min - REQUIRED 1.004 DISP Temperature: 23 +/-2 degC 1.004 DISP Humidity: <75% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 DISP This procedure requires a 4 1/2 digit DVM. 1.007 HEAD INITIAL SETUP 1.008 HEAD {mA INPUT} 1.009 DISP Set DISPLAY MODE to IN. 1.009 DISP Set INPUT to mA. 1.010 DISP Connect the 5520A to the UUT as follows: 1.010 DISP [32] 5520A AUX HI to UUT mA INPUT HI 1.010 DISP [32] 5520A AUX LO to UUT mA INPUT LO 1.011 5520 -80.00mA 0.01U 2W 2.001 5520 0.00mA 0.01U 2W 3.001 5520 80.00mA 0.01U 2W 4.001 HEAD {OHMS INPUT} 4.002 DISP Connect the 5520A to the UUT as follows: 4.002 DISP [32] 5520A NORMAL HI to UUT INPUT 1 4.002 DISP [32] 5520A NORMAL LO to UUT INPUT 2 4.003 DISP [32] 5520A AUX HI to UUT INPUT 3 4.003 DISP [32] 5520A AUX LO to UUT INPUT 4 4.004 DISP Set INPUT to OHMS. 4.004 DISP Set WIRES to 4. 4.005 HEAD OHMS INPUT: {200 Ohm Range} 4.006 5520 190.00Z 0.02U 4W 5.001 HEAD OHMS INPUT: {2000 Ohm Range} 5.002 5520 1900.0Z 0.2U 4W 6.001 HEAD {F deg INPUT} 6.002 DISP Set INPUT to deg F. 6.002 DISP Set RTD TYPE SELECT to PT 385. 6.003 5520 390.00Z S 4W 6.004 ASK- U 6.005 ACC 1560.4degF TOL 6.006 MEMI Enter UUT reading in degrees Fahrenheit: 6.007 MEME 6.008 MEMC degF 0.4U 7.001 ASK+ U 7.002 HEAD {OUTPUT MODE} 7.003 DISP Set INPUT to deg F. 7.003 DISP Set RTD TYPE SELECT to PT 385. 7.004 DISP Connect the 5520A to the UUT as follows: 7.004 DISP [32] 5520A AUX HI to UUT OUTPUT 1(3) 7.004 DISP [32] 5520A AUX LO to UUT OUTPUT 2(4) 7.004 DISP Connect the 34401A to the UUT as follows: 7.004 DISP [32] 34401A INPUT HI to UUT OUTPUT 1(3) 7.004 DISP [32] 34401A INPUT LO to UUT OUTPUT 2(4) 7.005 DISP Set DISPLAY MODE to OUT. 7.005 DISP Set OUTPUT to OHMS. 7.006 5520 1mA S 2W 7.007 MESS Adjust the COARSE and FINE OUTPUT controls on the 7.007 MESS UUT for a 34401A reading between 0.90V and 1.00V. 7.008 34401 0.95V S 2W 7.009 MEMI Enter measured value on UUT in volts. 7.010 MEME 7.011 MEMC V 0.001U 8.001 END